Rapid amplification of the RM-Yplex assay

Abuidrees, A, Alghafri, A and Hadi, S orcid iconORCID: 0000-0002-2994-3083 (2016) Rapid amplification of the RM-Yplex assay. Electrophoresis, 37 (21). pp. 2817-2821. ISSN 0173-0835

[thumbnail of Author Accepted Manuscript]
Preview
PDF (Author Accepted Manuscript) - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

231kB

Official URL: http://doi.org/10.1002/elps.201500549

Abstract

A multiplex PCR assay consisting of 13 Rapidly Mutating Y STR loci called RM-Yplex was previously developed. Platinum® Taq DNA polymerase was used to amplify the 13 Y STR loci in a single reaction at an amplification time of approximately 2.5 h. In order to shorten the process with reliable results, two DNA polymerases were tested with the multiplex. Phusion® Flash High Fidelity, TAKARA Z-taqTM, and Platinum® Taq DNA polymerases were investigated for conducting RM-Yplex assay at various PCR cycling conditions. Rapid, robust, and efficient amplification of all the markers within the multiplex were achieved. The amplification time was reduced from 2.5 h to less than 28 min with Phusion® Flash High Fidelity DNA polymerase using Veriti® PCR thermal cycler.