Sensor systems for bacterial reactors: A new flavin-phenol composite film for the in situ voltammetric measurement of pH

Casimero, Chamete, McConville, Aaron, Fearon, John-Joe, Lawrence, Clare Louise orcid iconORCID: 0000-0003-0170-0079, Taylor, Charlotte May, Smith, Robert B orcid iconORCID: 0000-0002-2829-5360 and Davis, James (2018) Sensor systems for bacterial reactors: A new flavin-phenol composite film for the in situ voltammetric measurement of pH. Analytica Chimica Acta, 1027 . pp. 1-8. ISSN 0003-2670

[thumbnail of Sensors - 1-s2.0-S0003267018305397-main.pdf] PDF - Published Version
Restricted to Repository staff only

1MB
[thumbnail of Author Accepted Manuscript]
Preview
PDF (Author Accepted Manuscript) - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

907kB

Official URL: https://doi.org/10.1016/j.aca.2018.04.053

Abstract

Monitoring pH within microbial reactors has become an important requirement across a host of applications ranging from the production of functional foods (probiotics) to biofuel cell systems. An inexpensive and scalable composite sensor capable of monitoring the pH within the demanding environments posed by microbial reactors has been developed. A custom designed flavin derivative bearing an electropolymerisable phenol monomer was used to create a redox film sensitive to pH but free from the interferences that can impede conventional pH systems. The film was integrated within a composite carbon-fibre-polymer laminate and was shown to exhibit Nernstian behaviour (55 mV/pH) with minimal drift and robust enough to operate within batch reactors.