Myocardial perforation associated with the use of the Gensini ventriculography catheter

Davis, Gershan orcid iconORCID: 0000-0001-9096-5495, Au, John and Roberts, David (1996) Myocardial perforation associated with the use of the Gensini ventriculography catheter. International Journal of Cardiology, 53 (2). pp. 103-106. ISSN 0167-5273

Full text not available from this repository.

Official URL: https://doi.org/10.1016/0167-5273(95)02519-7

Abstract

We describe myocardial perforation during left ventriculography using the Gensini catheter in two patients with aortic valve disease. In each case this occurred despite careful catheter positioning and prior 'test' injection of contrast agent. The use of the Gensini catheter for left ventriculography in this circumstance is seriously questioned.